Protein fluorophores, such as phycoerythrin (PE), allophycocyanin (APC), and PerCP, are bright fluorescent proteins used in flow cytometry to detect cellular markers. Tandem dyes, such as PE-Cy7 and APC-Cy7, combine a protein fluorophore with an acceptor dye, transferring excitation energy through FRET to produce fluorescence at longer wavelengths. These tandems expand the number of detectable channels in multicolor flow cytometry but require careful compensation due to spectral spillover and potential dye degradation.
Properties of Fluorescent Proteins
| PE | APC | PerCP |
|---|
| Laser | Blue (488 nm) and Yellow-Green (532/561 nm) | Red (633/640 nm) | Blue (488 nm) |
| Protein Ex/Em, filter | 565/574 nm, 575/26 | 651/660 nm, 670/14 | 477/678 nm, 670/30 |
| Brightness (ε × QY) | ε 1,960,000 M⁻¹cm⁻¹ × QY 0.82: the brightest of the three | ε 730,000 M⁻¹cm⁻¹ × QY 0.68: about a third of PE | ε 406,000 M⁻¹cm⁻¹ × QY 0.40: about a tenth of PE |
| Emission range | Orange (protein) to near-IR (tandems) | Far-red (protein) to near-IR (tandems) | Far-red (protein) to near-IR (tandems) |
| Best for | Low-abundance antigens and rare cell populations, on either the blue or the yellow-green laser | Red-laser channels clear of PE; CL-APC stays intact in storage and at the dilute concentrations that dissociate APC | Far-red to near-IR detection from the 488 nm laser alone, beside FITC and PE in single-laser panels |
Common Tandems and Their iFluor® Alternatives
| Instead of... | Upgrade to... | Detection Filter |
|---|
| PE-Texas Red | PE-iFluor® 594, PE-iFluor® 597 | 610/20 |
| PE-Cy5, PE-Alexa Fluor® 647 | PE-iFluor® 647 | 670/30 |
| PE-Alexa Fluor® 700 | PE-iFluor® 700 | 720/30 |
| PE-Cy7, PE-Alexa Fluor® 750 | PE-iFluor® 750 | 780/60 |
| APC-Alexa Fluor® 700 | APC-iFluor® 700 | 720/30 |
| APC-Cy7, APC-Alexa Fluor® 750 | APC-iFluor® 750, APC-iFluor® A7 | 780/60 |